81
iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Clog Prevention and Management
Using the Clog Detection Feature
By default, every new iQue Experiment has the Clog Detection feature enabled in the Protocol of every plate:
This is a software feature that works by monitoring the event rate during data acquisition. If the event rate fall below
a certain threshold for longer than expected, taking into account shakes and cleans which naturally lower the event
rate, then the clog detector triggers a warning:
Click
Yes
to stop the run immediately. Click
No
to dismiss the dialog and allow data acquisition to continue.
At this time, data acquisition does NOT stop unless the user chooses to end the run. If the clog detector warning
is going off frequently where there is no clog (false positives) the experiment protocol should be modified and the
Enable Clog Detection option unchecked.