Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Clog Prevention and Management
Signs and Symptoms of Clogging
In addition to the Clog Detection feature, there are several signs and symptoms of incipient clogging or other
cleanliness issues.
•
System appears to be operating normally but no events are being recorded.
•
Events are being recorded but, when data is analyzed, blocks of wells appear to be missing.
•
After clicking
Run
, the system takes well over a minute to begin acquiring data.
•
At the end of a run, or after clicking
Stop
, the system takes well over a minute to fully stop.
•
When running a sample from a tube, the sample is being consumed at an unusually fast rate and the
sheath pump is running abnormally fast.
Troubleshooting a Clog
The first step in troubleshooting a clog is to contact IntelliCyt tech support so that we may help assess the
seriousness and nature of the problem and recommend remediation steps. These may include performing standard
cleaning procedures such as the
Backflush
,
Unclog
,
Quick Clean
and
Long Clean
procedures.
Summary of Contents for iQue SCREENER
Page 1: ...iQue SCREENER hardware manual October 2014...
Page 2: ......
Page 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Page 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Page 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Page 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 110: ......