Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Changing the Probe and Tubing
Changing the Probe and Tubing
Overview
As the iQue runs, the main sample tubing is worn by the rollers of the peristaltic pump. To ensure proper sample
collection, IntelliCyt recommends that the probe and tubing assembly be changed after 15 hours of usage. This
article describes the proper way to perform this procedure.
Required Materials
Part Number
Description
90659
iQue
®
Probe & Tubing Assy for Gen 2 iQue
®
& iQue
®
HD - 5 Pk
90660
iQue
®
Probe & Tubing Assy for Gen 2 iQue
®
& iQue
®
HD - 10 Pk
90293
iQue
®
Probe & Tubing Assy for Gen 1 iQue
®
(8” Probe) - 5 Pk
90163
iQue
®
Probe & Tubing Assy for Gen 1 iQue
®
(8” Probe) - 10 Pk
Launch the Change Probe and Tubing Wizard
To begin changing the Probe and Tubing, click
Device > Change Probe and Tubing
...
Follow the Change Probe and Tubing wizard to complete the process. Some details may be slightly different
depending on which generation probe the instrument is equipped with, but the overall process is the same.