2. Prepare equipment for soaking by disassembling
removable parts and loosening connections.
3. Rinse equipment in warm water before placing
into bath.
4. Completely soak equipment in water/detergent
solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush.
NOTE
– If possible, use a disposable brush.
6. Rinse in warm purified water until all visible
substances and residual detergent are removed.
NOTE
– Make sure to give special attention to
hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels
and use medically compressed air if needed, to
dry channels, crevices and lumens.
8. Inspect the equipment to make sure there is no
visible organic debris or residue from cleaning
agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness
or excess liquid is detected, dry with a soft clean
towel.
• Medically compressed air can be used if needed.
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and
solutions cause damage to the devices.
• Channels and crevices found on these devices
require particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can
damage the surface of the equipment. Avoid
using hard water. Instead use purified water
unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on
these devices and is not recommended for
routine use.
• It is important to have Integra NeuroSpecialists
perform routine inspections (twice yearly is
recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH
Detergent
Method
1. Prepare equipment for cleaning by disassembling
removable parts and loosening connections.
2. Rinse equipment in warm water before placing
into washer.
3. Load device into the washer and place small parts
in container or tray inside the washer unit in order
to avoid losing small components.
NOTE
– Load devices carefully into washer in
order to avoid collision.
4. Follow the instructions listed below and set
washer machine to these exact parameters:
Phase
Time
(Min.)
Water
Tempera-
ture
Detergent and
Concentration
Pre-
wash 1
4:00
Cold water N/A
Enzyme
Wash
4:00
Hot water Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with
APA, Ratio 1:128)
Wash 1
10:00
60.0ºC
(140º F)
Neutral pH
detergent
(e.g. Renu-
Klenz™ , Steris,
Ratio 1:256)
Rinse 1
0:30
Hot water N/A
Thermal
Rinse**
2:00
82.2ºC
(180º F)
N/A
** Optional phase for disinfection of components
– minimum water temperature as indicated or per
worker manufacturer specifications for the thermal
rinse cycle.
NOTE
– Any deviation from this guideline could
result in damage to the equipment as well as
improper cleaning results.
Rinse with purified water. Do not perform if
parameters cannot be achieved.
Summary of Contents for MAYFIELD A1059
Page 2: ...This page is intentionally left blank...
Page 12: ...This page is intentionally left blank...
Page 13: ...Clamp cr nien MAYFIELD modifi A1059 Mode d emploi REF...
Page 22: ...Cette page est laiss e intentionnellement vide...
Page 23: ...Cette page est laiss e intentionnellement vide...
Page 24: ...Testiera modificata MAYFIELD A1059 Manuale di istruzioni REF...
Page 33: ...Questa pagina stata lasciata intenzionalmente in bianco...
Page 34: ...Modifizierte MAYFIELD Sch delklemme A1059 Gebrauchsanleitung REF...
Page 43: ...Diese Seite wurde absichtlich frei gelassen...
Page 44: ...Clamp Craneal MAYFIELD Modificado A1059 Manual de instrucciones REF...
Page 53: ...P gina dejada intencionalmente en blanco...
Page 54: ...Gemodificeerde MAYFIELD schedelklem A1059 Gebruikershandleiding REF...
Page 63: ...Deze pagina is bewust leeg gelaten...
Page 64: ...Modifierad MAYFIELD skallkl mma A1059 Instruktionshandbok REF...
Page 73: ...Denna sida har avsiktligen l mnats blank...
Page 74: ...Denna sida har avsiktligen l mnats blank...