17
5.3
T
HE
M
AINTENANCE
M
ENU
Selecting
Maintenance
from the Main Menu will display instrument procedures (see Section 6
Maintenance
Procedures
):
Start Up
Shut Down
Prime
Elute
Return
•
Selecting
Start Up
will take the user through a series of prompts for the daily Start Up procedure (see Section
6.2
Start Up Procedure
).
•
Selecting
Shut Down
will take the user through a series of prompts for the daily Shut Down procedure (see
Section 6.3
Shut Down Procedure
).
•
Prime
is for priming the elution fluid lines when a new elution fluid canister is installed in the unit. This
ensures that there is no air in the fluid path (see Section 6.4
Priming the Elution Fluid
).
•
Elute
performs a foam elution according to the current instrument settings. This can be used at any time to
test the elution settings (see Section 8
The Elute Function
).
•
Return
will lead back to the Main Menu.
5.4
T
HE
I
NSTRUMENT
S
ETTINGS
M
ENU
There are seven instrument settings. A description of each are listed below:
•
Beep volume
changes the volume of the beeps that occur at the end of concentration cycles and elutions.
The four options are
low
,
med.
,
high
, and
off
. Default = low
•
Power LED
turns the blue LED light in the Fluidics head on and off. Default = on
•
Run LED
turns the flashing red LED light in the Fluidics head on and off. Default = on
•
Elute LED
turns the flashing green LED light in the Fluidics head on and off. Default = on
•
Set Password
allows the user to set a password that must be entered before a custom protocol can be
created, modified, or deleted.
•
Protocol
allows the user
to set the instrument to standard (
Std
) or advanced (
Adv
) options. If set to
Std
, the
user will have access to two standard set points -
Valve Open
and
Pulse Count
. If set to
Adv,
the user will
have access to 11 advanced settings during Protocol Creation. Once set, the options are accessed in the
Protocol Setup menu under Create Protocol. Default = Std
•
Fluid Gauge
turns the Elution Fluid Volume gauge on and off. Default = on
•
Return
will return you to the Options menu.
FIGURE 5.3 MAINTENANCE DISPLAY