Copyright 2018, InnovaPrep LLC
Patented
https://www.innovaprep.com/patents
13
4
I
NSTALLATION OF THE
I
NSTRUMENT COMPONENTS
This Section will guide the user through setting up the instrument.
4.1
I
NSTALLING THE
P
OWER
S
UPPLY
Plug the power cord and power supply together, plug the power supply into the 12VDC port on the side of the unit, then
plug the power cord into an available power outlet. It is not necessary to select the correct voltage; the sensor accepts
line voltage of 100 to 240 VAC, 50-60 Hz, single phase. The connection is self-regulating.
WARNING: DO NOT USE ANY
POWER SUPPLY OTHER THAN THE ONE SUPPLIED WITH THE CP SELECT INSTRUMENT-DOING SO MAY CAUSE DAMAGE TO
THE INSTRUMENT AND VOID THE WARRANTY.
(NOTE: THE POWER SUPPLY FOR THE PREVIOUSLY RELEASED CP - 150
CONCENTRATING PIPETTE INSTRUMENT IS NOT COMPATIBLE WITH THE CP SELECT)
4.2
I
NSTALLING AND
R
EMOVING
T
HE
P
ERMEATE
L
INE
T
UBING
Included with the instrument is a Section of clear PVC tubing with a right angle quick connect fitting; this is the permeate
line tubing. To install the Permeate Line, first make sure that the metal button on the Permeate Port is pushed down on
the instrument. Then, insert the quick connect fitting on the Permeate Line into the
Permeate Port. When the fitting is fully seated the metal button will pop up, thus
locking it in place. Place the open end of the permeate line tubing into a waste
container of adequate volume relative to your sample volume.
To remove the Permeate Line, simply
press the metal button on the
Permeate Port and it will be released.
4.3
I
NSTALLING THE
S
TORAGE
F
LUID AND
F
LUID
A
DAPTER
A daily Shut-down Maintenance Procedure
is described in Section 6.3. The
procedure requires the use of the following items:
•
Storage Fluid (HC08558)
•
Reusable Storage Fluid Adapter (HC08560)
•
Reusable Maintenance Tip (HC08005)
These items are conveniently stowed on a bracket on the right side of the
instrument as pictured at right in
Figure 4.2
. Figure 4.4, below, provides a
closer view of the Storage Fluid and Storage Fluid Adapter.
To install the Storage Fluid Adapter, remove the orange rubber cap (save for later use-used for transport and shipping)
from the Canister Interface Port and clean the port using an Alcohol Prep Pad. Remove the sealed cap from the Storage
Fluid bottle, and install the metal sipper tube portion of the Fluids Adapter into the bottle. The metal portion of the
assembly can be adjusted up or down in the bung to reach the bottom of the fluid bottle by loosening the fitting located
on the top of the bung. The Bung can now clip onto the bracket as shown, with the Fluids Adapter tubing clipped into the
bracket behind the bottle. The Maintenance Tip also stows on the bracket in the 1
st
position. See Section 6 for detailed
instructions for use.
FIGURE 4.2 STORAGE FLUID BRACKET
FIGURE 4.1 PERMEATE LINE
INSTALLATION