Copyright 2018, InnovaPrep LLC
Patented
https://www.innovaprep.com/patents
34
12
C
HEMICAL
C
OMPATIBILITY
G
UIDE
When performing concentration cycles, it is important to know the sample matrix. This chart details how compatible the
Concentrating Pipette fluidics and the membrane material in the CPT are with certain chemicals.
R=Recommended
L= Limited Exposure
NR=Not Recommended
U=Unknown
0.1 µm
Polyethersulfone
CPTs
0.4 µm Polycarbonate
Track - Etched (PCTE)
CPTs
Polysulfone Hollow Fiber CPTs
(0.45 µm, 0.2 µm, 0.05 µm and
Ultrafiltration)
5% Acetic acid
R
R
R
25% Acetic acid
L
L
L
Acetic acid (glacial)
NR
NR
NR
Acetone
NR
NR
NR
Acetonitrile
NR
U
NR
0.1 N Ammonium hydroxide
R
R
R
Conc. Ammonium hydroxide
NR
NR
NR
Amyl acetate
NR
NR
NR
Amyl alcohol
NR
NR
NR
Aniline
NR
NR
NR
Benzene
NR
NR
NR
Butyl acetate
NR
NR
NR
Butyl alcohol
L
L
L
Carbon tetrachloride
NR
NR
NR
Chloroform
NR
NR
NR
Chromic acid
NR
NR
NR
Cresol
NR
NR
NR
Cyclohexanone
NR
NR
NR
Diacetone alcohol
NR
NR
NR
Dimethyl formamide
NR
NR
NR
Dimethyl sulfoxide
NR
NR
NR
Ethers
NR
NR
NR
Ethyl acetate
NR
NR
NR
Ethyl Alcohol
L
L
L
Fuels
U
U
U
50% Ethyl alcohol
L
L
L
95% Ethyl alcohol
NR
NR
NR
Ethylene dichloride
NR
NR
NR
Ethylene glycol
R
R
R
10% Formaldehyde
R
R
L
30% Formaldehyde
L
L
L
50% Formaldehyde
NR
NR
NR
Glycerin
R
R
R
Hexane
NR
NR
NR
5% Hydrochloric acid
L
L
L