1-9
For Safe Operation
1
Flat Panel Sensors Operation Manual 897N200608
Further Information for IEC 60601-1-2 (EN 60601-1-2)
1. Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the accompanying documents.
2. Portable and mobile RF communications equipment can affect medical electrical equipment.
3. Information regarding the cable affecting EMC is as follows.
Name
Connected Device
Maximum Length
General Specification
Network Cable
Between the RICMP
and the RICMC
20m (65.6 ft)
Cat5e or more,
UTP type and straight cable
Between the RICPB
and the Image processing
unit *
2m (6.56 ft)
* When the distance between the RICPB and the RICMC is 2 meters or more, install a commercially
available HUB. In that case, use a cable with a maximum length of 2 meters between the RICPB and HUB,
and a cable with a maximum length of 20 meters between HUB and the RICMC.
4. The use of accessories, transducers and cables other than those specified, with the exception
of transducers and cables sold by INNOLUX Corporation as replacement parts for internal
components, may result in increased emissions or decreased immunity of the RIC.
5. The RIC should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the RIC should be observed to verify normal operation in
the configuration in which it will be used.
6. Basic performance of the equipment and the system
After image data are acquired from the flat panel sensor, data correction is performed by the control
cabinet, and the image is saved in and displayed on the image processing unit.
7. Test items (Tables 1 to 4)
Table 1
Guidance and manufacturer’s declaration - electromagnetic emissions
The RIC is intended for use in the electromagnetic environment specified below.
The customer or the user of the RIC should assure that they are used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The RIC uses RF energy only for their internal function.
Therefore, their RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The RIC is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies