A-1
Appendix
A Specifications
Flat Panel Sensors Operation Manual 897N200608
Appendix A
Specifications
A.1 Specifications
Specifications of the flat panel sensor are shown below.
A.1.1 Reduced Equivalent (Flat Panel Sensor)
Peak reduced equivalent on the front panel of the flat panel sensor: 0.5 mmAl
A.1.2 Power Supply Conditions
Rated voltage : 6-12V
Input current
: 2.73A
A.1.3 Environmental Conditions
(1) Operating Conditions
Temperature
: 15°C (15%RH) - 30°C (80%RH)
Humidity
: 15%RH (15°C) - 80%RH (30°C) (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
(2) Non-operating Conditions
(Environmental conditions under which power can be supplied)
Temperature
: 5°C - 35°C
Humidity
: 10%RH - 80%RH (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
(3) Storage Conditions
Temperature
: -30°C - 50°C
Humidity
: 10%RH - 90%RH (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
CAUTIONS
When the flat panel sensor is used in high temperature condition for long period of time, it may
cause image artifacts and/or failure of the device.