Maintenance and Inspection
Flat Panel Sensors Operation Manual 897N200608
Maintenance and Inspection
1 Maintenance and Inspection Items Assigned to Specified Dealer
For periodical inspection of the equipment and necessary arrangements, consult our official dealer
or local representative.
Periodical Maintenance
Make sure that the periodical maintenance and inspection assigned to our official dealer or
INNOLUX Representative are performed as specified.
Maintenance and Inspection Items Assigned to Specified Dealer
Periodical Maintenance and Inspection Items
Period
Checking of the image
Every year
Checking of the operation record by referring to the error log
Every year
Checking of the flat panel sensor
Every 2 years
* It is recommended that the battery pack be replaced, if the battery storage capacity becomes
lower than 60%.
The battery pack should be replaced when the operable time is less than the following.
●
RIC 35G/RIC 43G/RIC 35C/RIC 43C/RIC 24C :
100 minutes (1 hour and 40 minutes)
* Refer to the operable time displayed on the image processing unit when the battery pack is fully
charged and no exposure menu is registered.
* Depending on the usage environment, etc., the displayed time is slightly different from the
actual operable time.
(Reference) Time for battery pack replacement
Battery storage capacity [%]
Operable time [Minute]
120
100
80
60
60
40
20
0
0
30
60
90
120
150
210
240
180
Time for replacement:
Less than 100 minutes
(1 hour and 40 minutes)
The cycles of periodical maintenance and inspection and of parts replacement differ depending on
the usage and the daily operation time.