6.4
Trigger & Alarms
•
>
The trigger levels, the volume of the loudspeaker and the units of leak rates and
pressures can be set in this menu
.
6.4.1
Trigger Level 1
•
>
>
The value of the first trigger level can be set.
See Chapter
for the description of the entry.
Trigger 1 and Trigger 2 are programmable leak rate levels. If these levels are
exceeded some outputs of the UL5000 will react, especially if Trigger 1 is exceeded.
Display
In the status line of the display the signs for Trigger 1 and Trigger 2 are displayed
inverted if the leak rate exceeds (becomes higher than) the programmed value.
Relay Output
The relays on the I/O interface
/2 will switch.
See Chapter
for further details.
Alarm/Loudspeaker
Trigger level 1 defines at which level the various alarm types react (
Menu
Key No.
Name
Description
1
Back
Return to the main menu.
2
Definition of Trigger level 1. Please refer to Chapter
3
Definition of Trigger level 2. Please refer to Chapter
4
Please refer to Chapter
5
Selection of leak rate and pressure units. Please refer to
Chapter
6
Not used in this menu.
7
Please refer to Chapter
8
Choice of different alarm types. Please refer to Chapter