Description of the Menu
51
iina
74
e 0
6
.f
m
te
chn
ica
l ha
nd
boo
k
(1
40
8)
6.3
Mode
•
>
enables the submenu to select the different working modes.
Fig. 6-4: The
Menu
Key No.
Name
Description
1
Cancel
Return to the main menu without any changes
of the present settings.
7
Vacuum
The normal vacuum mode will be used.
Vacuum mode with the possibility to use
HYDRO•S.
HYDRO•S improves your time to test. Water
vapor desorption in a test object generates an
unwanted background signal. HYDRO•S
separates and eliminates this signal from the
helium signal, so testing can begin sooner.
After changing to this mode, a new calibration
necessary. Please refer to Chapter
,
HYDRO•S automatic ON.
5
Not used in this menu.
3
Sniff
The sniff mode will be used. Please refer to
Chapter
8
OK
Save the settings and return to the previous
menu.