Description of the Menu
55
iina
74
e 0
6
.f
m
te
chn
ica
l ha
nd
boo
k
(1
40
8)
6.4.6
Audio alarm type
•
>
>
The trigger of the audio alarm can be switched on or off.
6.4.6.1
Pinpoint
The tone of the acoustical signal changes its frequency only in a LR-window
which ranges from one decade below the Trigger level 1 up to one decade above
the Trigger level 1. Below the window the tone is constantly
low
, above the window
it is constantly high.
Example: The Trigger level 1 is 4×10
-7
mbar l/s. So the window where the tone
changes reaches from 4×10
-8
mbar l/s up to 4×10
-6
mbar l/s.
6.4.6.2
Leak rate prop.
The frequency of the accoustic output is proportional to the reading on the bargraph
display. The frequency ranges from 200 Hz to 1300 Hz. See Chapter
for the definition of the number of decades.
Softkey 2:
See Chapter
Softkey 3:
The sound will be proportional to the leak rate signal.
See Chapter
Softkey 6:
See Chapter
Softkey 7:
An alarm sounds when the trigger 1 is exceeded.
See Chapter
Fig. 6-6: Pinpoint