Hydrogen leak detector Sensistor ISH2000 HySpeed
-
Operating Instructions
38
OP-Sensistor ISH2000 HySpeed-EN-201203
EN
* entered as integer in 100’s of seconds, 1= 0.01s, 100 = 1 s, 60000 = 600s.
Transfer of parameters
Send parameters one by one; first send the specific header (for example “A”), then
wait until the RS232 transmitter unit has sent the data (approximately 20 ms), and
then send the data (for example “1.00E+01”). Data string must be ended with a
carriage return character, chr13 (dec).
Example:
“CPPM” or “C PPM”, Carriage Return (chr 13). This sets Reject Level Rate
Unit to “PPM”.
Parameters can be sent in any order you like:
•
If your data was received and correct, Sensistor ISH2000 HySpeed immediately
echoes (sends back) the data.
•
If you send a non existing header you will not receive anything.
•
If the data could not be converted in the Sensistor ISH2000 HySpeed you will
receive the string “CoEr”, (Conversion Error).
Note:
Remember to use capitals for the header.
APC driver installation
APC drivers are installed in the detector from a PC. All active probes need a driver to
be installed before they can be used.
For driver installation you will need the following:
•
APC Driver software.
•
File transfer cable.
•
PC computer with Windows 95 or later.
Timer A
D
nnn*
Timer B
E
nnn*
Timer C
F
nnn*
Timer D
G
nnn*
Purge Level
H
n.nnE+nn
Reference Value
I
n.nnE+nn
Reference Unit
J
Text string (max 12 characters)
Reference Unit (Program settings)
JUx
x=1 to 8, 1=PPM, 2=CC/S
Parameter
Header
Data