Set Date
Real Time Clock date set as YY-MM-DD. Clock runs even when detector is
disconnected from the power supply.
Printer Port
The Sensistor ISH2000 HySpeed is equipped with a serial (RS232) printer port. See
“Printer port” on page 33.
Info
Contains information about software versions, Serial number, and Internet contact
information.
8.10 Service Settings
The Service Mode is reached by starting the instrument and at the same time hold
the right button down on the panel. After start a new main menu called Service
Settings will appear.
Show Password
Shows the chosen password in case the customer has forgotten the password.
Contact INFICON AB to have the code sent to you. See the web address under the
section-Info.
Instrument system Reset
Resets all parameters into the instrument to default settings. Contact INFICON AB to
have the code sent to you. See the web address under the section-Info.
Probe System Reset
Resets all parameters into the probe to default settings. Contact INFICON AB to have
the code sent to you. See the web address under the section-Info.
Detector Signal Level
The
Detector Signal Level
is the level which the sensor is considered to have recovered
from the last gas signal. It decides when the
DET_SIGNAL
output will come on. this
signal can be used to block the start of a calibration or new test cycle in semi and
fully automatic testers.
IF
DET_SIGNAL
is high then this means that the sensor has detected hydrogen and
has not yet recovered.
Detector Signal Level
can be adjusted in the
Service Settings menu
. You can increase
Detector Signal Level
if you have many small disturbing signals. A high setting of
Detector Signal Level
gives better tolerance to “noise” gas signals at the expense of
accuracy. A low setting gives best accuracy but lower tolerance for “noise” gas
signals. The
Detector Signal Level
is set as 1 to 100% of the
Reject Level
. Default is
20%.
NOTICE!
Increasing the
Detector Signal Level
may give poorer accuracy.