Hydrogen leak detector Sensistor ISH2000 HySpeed
-
Operating Instructions
28
OP-Sensistor ISH2000 HySpeed-EN-201203
EN
set brightness if any of the display buttons are pressed, if a gas signal is detected or
an instrument error is detected.
8.9 General settings
This part describes the general settings of the Sensistor ISH2000 HySpeed.
Language
The Sensistor ISH2000 HySpeed user interface contains the following languages:
•
English
•
French
•
German
•
Italian
•
Spanish
•
Swedish
Probe Button
This is for setting the different functions with the probe button. These functions are
as follows:
•
Toggle Mode-makes it possible to switch between Analysis mode and Detection
mode.
•
Zero detection signal-in Analysis mode and Detection mode.
•
Measure/Print-makes it possible to initiate sample cycles or send the values from
the hand probe measurement to the printer port.
•
Probe Lamp-makes it possible to turn on and off the Probe Lamp.
Probe Lamp
Makes it possible to have the Probe Lamp on even if the other Probe Button
function is chosen.
Change Password
The user password is a text string (max 12 alphanumerical characters) used to lock
critical parameters. Setting password to an empty string (no characters) means that
no password is needed to modify the critical parameters. The default is no password
(“”).
Contact INFICON AB if you have lost your user password. If the
Password Protected
Calibration
parameter is set to
ON
you will be prompted for a password when starting
a calibration.
Note:
Setting
Password Protected Calibration
to
ON
has no effect if no password is set.
Note:
APC controlled calibration can be started from the bus in both cases.
Audio Base Frequency
This sets the lowest audio base frequency tone in Analysis and Detection Mode.
Set Clock
Real time set as hh:mm:ss. Hours and minutes can be adjusted. Seconds will
automatically be set to 00 when hours and minutes have been set. Clock runs even
when detector is disconnected from the power supply.