Hydrogen leak detector Sensistor ISH2000 HySpeed
-
Operating Instructions
40
OP-Sensistor ISH2000 HySpeed-EN-201203
EN
Fig 8-7. Fixing the mounting brackets.
See “Sensistor ISH2000 HySpeed specifications” on page 43 for electrical
connections.
8.16 Default parameters
Table 8-14. Range and default settings of all Sensistor ISH2000 HySpeed parameters.
4.
Hold the panel in place while fixing the mounting
brackets to the detector (See Figure 8-7).
5.
Align the detector horizontally and lock it by
tightening the 4 set screws.
6.
Lock the set screws with the locking nuts.
7.
Connect ground terminal to protective ground of
cabinet.
Step
Action
Sensistor ISH2000 HySpeed photo of
brackets
Bracket to panel screw
Bracket to unit
Bracket
Parameter
Range
Default
Analysis Unit
Several Choices
PPM
APC Time A
0.0 - 6000.0 s
1.0 s
APC Time B
0.0 - 6000.0 s
0 s
APC Time C
0.0 - 6000.0 s
0 s
APC Time D
0.0 - 6000.0 s
0 s
Audio Base Frequency
Several Choices
400 HZ
Auto Range
ON/OFF
ON
Audio Ready Pulse
ON/OFF
ON
Audio Threshold
(Detection)
0 - 100%
0%
Audio Threshold
(Analysis)
0 - 100%
4%
Brightness
0 - 21
21
Calibration Time
Min Calibra. Time-30 s
10 s
Calibration Tolerance
0-100%
25%