Problem
Possible Cause
Possible Solution
Cannot configure level or other
parameters using the
Configure
button on the
Sensors
tab. The
Sensor Setup
screen is shown, but
the
Configure
... button is dim.
The instrument is actively polling
(continually updating real-time readings)
in the
Home
tab.
The instrument has an active log—a log
that is Ready, Pending, Running, or
Suspended as shown in the
Status
column of the
Logging
tab. Only one
active log can reside in the device at a
time.
Return to the
Home
tab and stop real-time
readings by clicking the
Play
button.
Stop or delete the log if possible. Alternatively,
configure parameters after the log is complete.