
1-800-446-7488
39
www.in-situ.com
Install the Desiccant
Vented cable must be protected with a desiccant pack that is properly sized for site
conditions.
Desiccant
Desiccant protects cables, connections, and internal components from condensation, which can cause irreparable damage
and loss of data. Indicating desiccant changes from blue to pink as it becomes saturated with moisture.
Desiccant stages (from left)
New, nearly expired (replace now), expired
It is
extremely important
to use a properly-sized desiccant for your deployment and to
change desiccant often
. Desiccant
should be changed before the entire volume has turned pink, and you should use enough desiccant to effectively keep your
equipment dry until your next scheduled maintenance. Desiccant longevity is dependent on site conditions and can vary from
one site to the next.
Installation Guidelines
l
Never let the instrument fall freely down a well. Doing so will damage the sensor.
l
After you have installed the instrument, verify the water level reading. Move the instrument and take another
reading to ensure that the instrument shows reasonable change. The instrument could be wedged against the well
casing with a loop of cable hanging below. An instrument in such a position could become dislodged and move
while it is logging data, which would record a false change in the water level.
l
For accurate measurements, the instrument should remain immobile while it is logging data.
l
Make sure that the uphole cable end is protected. The vented cable must be protected with a desiccant, and the
non-vented cable must be protected with a dust cap. The uphole cable end must be positioned above the highest
anticipated water level. Avoid placing this end in a location that might flood.
l
Do not deploy instruments in such a way that ice may form on or near the sensor or cable connections. Ice
formation is a powerful expansive force that can over-pressurize the sensor or otherwise cause damage. Damage
associated with ice formation is not covered by the instrument warranty.
l
Do not allow vented cable to bend enough to obstruct the internal vent tube. The recommended bend radius is 13.5
mm (0.54 in), which is twice the cable diameter.
Summary of Contents for Level TROLL 400
Page 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Page 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Page 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Page 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Page 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Page 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Page 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Page 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Page 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Page 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...