1-800-446-7488
57
www.in-situ.com
Stopping a Log
A log can be manually stopped at any time, even if a stop time has been previously scheduled. If you did not specify a stop
condition when you defined the log, the log will run until the instrument is out of memory or battery power, or until you
manually stop it.
A log that has been stopped cannot be resumed. If you intend to resume a log later, you
should
suspend a log
rather than stop it.
1. To manually stop a log, the instrument must be connected to the software.
2. Select the
Logging
tab
.
3. Select the running log you intend to stop.
4. Click the
Stop Log
button
.
Downloading Data to a PC
This procedure copies the data log from the instrument to a PC. It does not remove the data log from the instrument. After a
log is downloaded, it can be exported to a CSV file format that can be used by spreadsheet programs. The time shown in the
log name is the time the log was downloaded.
1. With an instrument connected, select the
Logging
tab
.
2. Select the log you intend to download.
3. Choose a Running, Suspended, Stopped, or Deleted log.
4. Click the
Download
button.
5. In the next screen, select one of the three download options.
l
All data
l
New data (data logged since the last download)
l
Time interval to download
New data is downloaded by default to a new log file. To append new data to the last download of
this log, be sure the option "Append logs on download" is selected in the
General Settings
dialog (
Preferences
>
General Settings
).
2. The log is copied to the connected PC into your Win-Situ working directory folder. View or change the working
directory using
File
>
Settings
.
3. At the end of the download, Win-Situ gives you the option of viewing the data.
l
Select
Yes
and the log is displayed in the
Data
screen.
l
Select
No
and the
Logging
screen appears. You can view the data at any time by selecting it in the
Data
tab.
Viewing Logged Data
1. To view the data stored in the instrument, you must first download the data. A connection to an instrument is not
needed after the data log has been downloaded.
2. Select the
Data
tab
.
Summary of Contents for Level TROLL 400
Page 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Page 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Page 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Page 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Page 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Page 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Page 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Page 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Page 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Page 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...