Cleaning and Maintenance
Overview
It is important for users to perform scheduled maintenance on their instruments to sustain the accuracy and longevity of the
probes and cables. The frequency of this maintenance depends on the characteristics of the deployment site, including
humidity levels and the degree of fouling.
Users should be aware of the conditions at their deployment sites and develop appropriate maintenance schedules to replace
desiccant, clean the instruments, and send in the instruments for factory calibration. Users should check instruments often
during the first portion of the deployment to determine the frequency of maintenance. General maintenance should be
performed as often as possible. Factory maintenance and calibration should be performed every 12 to 18 months.
Operating Considerations
The instrument has been designed to withstand harsh field conditions. However, as with any electronic instrument, it can be
permanently damaged if used outside its operating specifications.
Temperature
Review the instrument specifications to determine the operating range. Do not deploy instruments in such a way that ice may
form on or near the sensors or cable connections. Ice formation is a powerful expansive force that can over-pressurize the
sensor or otherwise cause damage that is not covered by the warranty.
Pressure Range
The instrument can withstand pressures of up to two times (2X) the rated range of the pressure sensor without damage,
although it may not read correctly at such pressure. If the pressure range is exceeded by 3X, the sensor will be destroyed.
Batteries
Internal batteries in the instrument are not user-replaceable. The approximate percentage remaining is displayed in Win-Situ
software or the Vu-Situ mobile app when the instrument is connected.
If batteries are completely exhausted, external power and battery pack options are available. See page 24.
Summary of Contents for Level TROLL 400
Page 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Page 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Page 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Page 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Page 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Page 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Page 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Page 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Page 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Page 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...