Setup Manual
Page 43
D.1
White Balance
Picture 35: Setup menu, start screen
The first menu item of the setup menus is the
White Balance
screen.
Whenever it is necessary to perform a White Balance calibration, the touchscreen shows
how to position the reference target for optimal calibration.
Place the reference target
WT36A-WA-01-A
as shown. The reference target is delivered
with the scanner.
Touch the
Calibrate
button.
During the calibration sequence the reference target will be transported forward and
reverse.
The calibration sequence will take approximately 15 seconds.
At the end of the calibration sequence, the results will be displayed on the touchscreen
and the reference target will be parked at the document table.