GENERAL INFORMATION
1-9
Modulation ALC:
Levels the applied external modulation over the range 0.75 to 1.25 V
RMS. High and low indicators in display indicate when the input is
outside leveling range.
Sweep mode
Any or all of the sources may be set to a carrier frequency sweep mode.
The sweeps are defined by entry of the start, stop and step frequencies.
A common step time can be set from 50 ms to 10 s per step. A
common trigger input at a rear-panel BNC connector may be used to
trigger a step or a complete sweep. Sweep can also be set to
continuous.
Frequency standard
The carrier frequency and internal modulation frequency are
synthesized from either an internal reference oscillator or an external
reference.
Internal standard:
10 MHz OCXO.
Aging rate:
±
2.5 in 10
7
per year.
±
5 in 10
9
per day after 2 months’ continuous use.
Temperature stability:
Better than
±
7 in 10
8
over the temperature range 0 to 50
°
C.
Warm
-
up time:
Within 2 in 10
7
of final frequency 10 minutes after switch on at a
temperature of 20
°
C.
External input:
Requires an input of 220 mV RMS to 1.8 V RMS into 1 k
Ω
on rear-panel
BNC connector. Input frequency can be 1 MHz or 10 MHz.
External output:
Rear-panel BNC socket provides an output of 10 MHz at a nominal level
of 2 V pk-pk into 50
Ω.
Remote control
GPIB:
All standard signal generator functions except the supply switch are
remotely programmable.
Capabilities:
Complies with the following subsets as defined in IEEE Std 488.1: SH1,
AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2.
RS-232:
All standard signal generator functions except the supply switch are
remotely programmable.
Connector is 9-way male D-type, baud rate 300 to 9600 bits per second.
Handshake hardware is DTR, RTS, CTS and DSR, and software is
XON and XOFF.
Electrical interface is to EIA-232-D.
Electromagnetic compatibility
Conforms with the protection requirements of the EEC Council Directive
89/336/EEC
Conforms with the limits specified in the following standards:
IEC/EN 61326-1 : 1997, RF Emission Class B,
Immunity Table 1, Performance Criterion B
Safety
Conforms with the requirements of EEC Council Directive 73/23/EEC
(as amended) and the product safety standard IEC/EN 61010-1 : 2001 +
C1 : 2002 + C2 : 2003 for Class 1 portable equipment, for use in a
Pollution Degree 2 environment. The instrument is designed to operate
from an Installation Category 2 supply.
Rated range of use
(over which full specification is met unless otherwise indicated).
Temperature
0
°
C to +55
°
C.
Humidity
Up to 93% at 40
°
C.
Altitude
Up to 3050 m (10 000 ft).
Conditions of storage and transport
Temperature
−
40
°
C to +71
°
C.
Humidity
Up to 93% at 40
°
C.
Altitude
Up to 4600 m (15 000 ft).
Power requirements
Voltage: 100 to 240 V~ (limit 90 to 264 V~)
Frequency:
50 to 60 Hz (limit 45 to 66 Hz)
Power consumption:
250 VA maximum.
Calibration interval
Recommended 2 years. Realignment can be accomplished by GPIB
control or from the front panel There are no mechanical adjustments