![IFR 2026Q Operating Manual Download Page 89](http://html1.mh-extra.com/html/ifr/2026q/2026q_operating-manual_3876146089.webp)
MEMORY
3-47
Memory protection menu
(4) To change the write protection, the instrument must be unlocked to Level 1 (see ‘Protection
locking and unlocking’ on page 3-37). Subsequently pressing
[Memory Protect]
causes the
Memory Write Protection Menu
similar to that shown in Fig. 3-45 below to be displayed.
B3935
Fig. 3-45 Memory write protection menu
(5) Select the type of memory stores to be changed by selecting between
[Full Stores], [Carrier
Stores]
and
[RAM Stores].
An unprotected store is indicated by a dash (
−
), a protected store
is indicated by the letter
p
.
(6) To change the protection for a memory block enter the
[Block Start]
and
[Block Stop]
numbers. For a single store make both numbers the same.
(7) Select
[Protect]
or
[Unprot.]
as required. This action is acknowledged by the message
** Protected **
or
** Unprotected **
appearing respectively.
(8) Press
[EXIT]
to return to the
Memory Store Menu
.
Note that at power-on the volatile RAM stores are unprotected to allow immediate use.
Memory recall
There are three types of recall: carrier, full and RAM. Both carrier and full stores are non-volatile.
The contents of the RAM store are lost when the instrument is switched off.
Carrier recall
The non-volatile carrier frequency store has 100 locations numbered 0 to 99 for carrier frequency
only. These can be recalled and used in conjunction with full recall to apply a set of test conditions
to a range of frequencies.
Full recall
The non-volatile store has 100 locations numbered 0 to 99 for the storage of instrument settings.
These stores may be recalled and used to reset the instrument’s parameters to those which affect
the RF output: carrier frequency, RF level, modulations in use, on/off and source information and
the two modulation oscillator frequencies in use.
RAM recall
The volatile RAM store has 100 locations numbered 0 to 99 for the full storage of instrument
settings. The parameters that are recalled are the same as those for full recall.