UTILITIES
3-36
Remote control selection
(menu 1)
The remote mode of operation is selected as follows:
(1) Select
[Remote Control]
to display the
Remote Control Utility
. This shows the currently
selected remote mode (see Fig. 3-29 below).
B3920
Fig. 3-29 Remote control utility
(2) Press
[GPIB/RS232]
to toggle between
GPIB
and
RS232
.
GPIB mode
(3) Press
[GPIB Address]
and enter the address, which must be unique on the system to the
instrument and within the range 0 to 30.
RS-232 mode
(3) Press
[Data Bits]
to toggle between
7
and
8
data bits.
(4) Press
[Stop Bits]
to toggle between
1
and
2
stop bits.
(5) Press
[Parity]
to cycle through the selections
None, Odd, Even
.
(6) Press
[Hardware Hndshk]
and
[XON/XOFF Hndshk]
to select any combination between both
OFF
to both
ON
.
(7) Press
[Baud Rate]
and set the baud rate in the range 300 to 9600 bit/s. But note that if the
entered value is not one of the following standard settings, the next higher value will be
selected by default.
300
600
1200
2400
4800
9600
(8) Press
[EXIT]
to return to the
Utilities Selection Menu 1.