9.5.2 Configure standard display
►
Select [SELd] and set the standard unit of measurement:
-[FLOW] = the current volumetric flow value in the standard unit of
measurement is displayed�
-[TOTL] = the current meter reading in Nm
3
or Nm
3
x 10
3
is displayed�
-[TEMP] = the current medium temperature in °C is displayed�
►
Select [diS] and set the update rate and orientation of the display:
-[d1] = update of the measured values every 50 ms�
-[d2] = update of the measured values every 200 ms�
-[d3] = update of the measured values every 600 ms�
-[rd1], [rd2], [rd3] = display as for d1, d2, d3; rotated by 180°�
-[OFF ] = the display is switched off in the operating mode�
9.5.3 Set output status in fault condition
►
Select [FOU1] and set the value:
-[On] = output 1 switches ON in case of a fault�
-[OFF] = output 1 switches OFF in case of a fault�
-[OU] = output 1 switches irrespective of the fault as defined with the
parameters�
►
Select [FOU2] and set the value:
-[On] = output 2 switches ON in case of a fault, the analogue signal
goes to the upper final value�
-[OFF] = output 2 switches OFF in case of a fault, the analogue signal
goes to the lower final value�
-[OU] = output 2 switches irrespective of the fault as defined with the
parameters� The analogue signal corresponds to the measured value�
9.5.4 Set measured value damping
►
Select [dAP] and set the damping constant in seconds (t value 63 %)�
9.5.5 Set low flow cut-off
►
Select [LFC] and set the limit value
Flows below the LFC value have no effect on the display and output signals�
9.5.6 Calibrate curve of measured values
►
Select [CGA] and set a percentage between 60 and 140 (100 = factory
setting)�