ENG
Resetting menu language
23
RESETTING MENU LANGUAGE
If the measuring instrument is set to the incorrect lan-
guage, you may not be able to operate it.
You can reset the menu language without under-
standing the menu entries.
How to reset the menu language
1
Press
[F4]
until the instrument switches off.
2
Press
[F4]
to switch the measuring instrument on.
3
Press
[F1]
,
[F2]
and
[F3]
in succession while the
measuring instrument switches on.
4
Set the menu language.
How to define the basic settings, page 7
RESETTING MEASURING INSTRUMENT
You can reset the measuring instrument. Resetting
has the following effects:
•
All settings in the
CHANNELS
menu are reset.
•
All measurement series in the
MEMORY
menu
are deleted.
•
The settings in the
DEVICE (1/2)
menu are reset:
• Language
• Unit
• Disp.rate
The settings in the
CALIBRATION
,
TIME
and
DATE
menus remain unchanged.
When you switch the instrument on, you must make
the basic settings.
How to reset the measuring instrument
1
If possible, save all stored measurement series
(e.g. with HYDROlink6).
2
Press
[F4]
for 2 seconds to switch off the measur-
ing instrument.
3
Press
[F1]
,
[F2]
and
[F3]
in succession while the
measuring instrument switches off.
The measuring instrument is reset and switches
off.