Instrument configuration
22
MultiPanel 2025
O
perati
ng Instruct
ions
• Version
1.0
Reminder of the next calibration
You can determine when the measuring instrument
displays a reminder. When switching on, the measur-
ing instrument checks when the measuring instru-
ment was calibrated and which calibration interval is
set.
If the time to the next calibration is less than the set
reminder period, the measuring instrument displays
the following information after the next switch-on:
If the date of the next calibration has been exceeded,
the measuring instrument displays the following infor-
mation after the next switch-on:
You must confirm the information with
[F4]
in order to
use the measuring instrument.
How to change the reminder period
1
Open the
DEVICE
menu and use
[F1]
to select
the
Calibration
entry.
How to open the
DEVICE
menu, page 20
2
Press
[MENU]
.
3
Set the reminder period:
•
Use
[F1]
to select the
remind
entry and press
[MENU]
.
•
Press
[F1]
or
[F2]
to change the value.
•
Use
[MENU]
to save your entry.
4
Press
[F4]
to exit the
CALIBRATION
menu.
Time
Read the following section for information on the
CLOCK MENU:
:
Date
Read the following section for information on the
DATE MENU
:
Error code
Information text
Date of the next calibration
Error code
Information text
Date of last calibration