A-26
A4
C.1 interface_name
The interface_name procedure is used to read/modify the current VXI-11 logi-
cal interface name.
struct Int_Name_Parms {
unsigned int action;
unsigned int length;
opaque
name<>;
};
struct Int_Name_Resp {
unsigned int error;
unsigned int length;
opaque
name<>;
};
Int_Name_Resp interface_name (Int_Name_Parms) = 1;
The action value determines whether the client wishes to execute a read of the
current setting, or a modification of the current value.
action = 0 = read of current value
action = 1 = modify current value
If the action value is other than 0 or 1, then an error value of 5 is returned.
If the RPC message is of insufficient length to satisfy the required length, an
error value of 1 is returned.
The name string must be a NULL terminated string with a 32-byte maximum
length (exclusive of the NULL). An error of 5 is returned and the Interface
Name is unchanged if the name field exceeds 32-bytes.
The returned Int_Name_Resp structure will always contain the current Interface
Name, irrespective of the error value.
error Meaning
0
No error
1
Syntax error
5
Parameter error
Summary of Contents for 9099
Page 20: ...1 16 1 Figure 1 1 9009 Outline Dimensions ...
Page 24: ...1 20 1 Figure 1 3 9099 Certificate of Compliance ...
Page 54: ...3 8 3 Figure 3 1 Status Reporting Structure Flash Configuration Reset ...
Page 94: ...3 48 3 This page left intentionally blank ...
Page 100: ...4 4 6 This page is left intentionally blank ...