2-2
2
2.4
INSTALLATION GUIDES
2.4.1 9009 Installation Guide
The following steps should be used as to setup and use the 9009.
1. Review Section 2.9 to select or design the serial cable to the Modbus
RTU Slave Device. Change the jumper positions on the 9009 as
directed in Section 2.10 to match the selected signal type.
2. Select a convenient location to mount the 9009. The preferred loca-
tion is flush against the rear panel so that its Ethernet, GPIB and USB
connectors protrude through the rear panel and provide good RFI/
EMI suppression. Use the cutouts and mounting dimensions shown in
Figure 2-1 for rear panel mounting. Panels thicker than 0.050 inches
will need counterboring to keep the lockstuds in their correct positions.
Provide a 0.1 inch clearance between the 9009 and any metal surface.
Mount away from any heat producing surface and high frequency, high
current cables.
.000
.062
.218
.320
.163
.225
.283
.000
.200
1.730
3.300
4.125
4.900
5.300
5.500
DETAIL 'A'
DETAIL 'B'
2X ÿ.150
ÿ.100
DETAIL 'C'
.000
.010
1.575
1.842
.625
.312
DETAIL 'A'
CL
.150 DIA THRU
2 PL
DETAIL 'B'
.500
.450
.656
.580
DETAIL 'C'
Figure 2-1 9009 Rear Panel Cutouts
Summary of Contents for 9099
Page 20: ...1 16 1 Figure 1 1 9009 Outline Dimensions ...
Page 24: ...1 20 1 Figure 1 3 9099 Certificate of Compliance ...
Page 54: ...3 8 3 Figure 3 1 Status Reporting Structure Flash Configuration Reset ...
Page 94: ...3 48 3 This page left intentionally blank ...
Page 100: ...4 4 6 This page is left intentionally blank ...