3-33
3
You can now run the VISA Interactive Control Panel.
1.
Select the INSTR TCP/IP Resource you just entered.
2.
Click the Open VISA Session button.
3.
Select the BASIC I/O tab.
4.
Select the Write tab. *IDN? is prefilled in the text box but you can
change it to any command. Press Execute to send the command to
the GPIB device.
5.
Select the Read tab. Press Execute to read the device’s response. Only
execute the read function if you are expecting a reply from the device.
Otherwise you will get a timeout error.
6.
The remaining tabs let you send a Device Trigger or Device Clear to
the device and Serial Poll the device.
3.7.7 Direct Control with TCP/IP
The 9099's WebServer communicates thru port 80. If you write a program to
open a socket to port 80 on the 9099 and send a single line containing the pa-
rameter via the socket, it would be the same as having a browser communicate
with the 9099. You can create a small program to visit, examine or control
the html page variables. Refer to the html pages supplied with your 9099 for
the variables names and to Application Bulletin AB80-17 for more detailed
information on direct control with port 80. An example is:
GET/cgi?%input1%\r\n
'reads F4 temperature