MediaClave overview
INTEGRA Biosciences
Page 58
7.6.4
Interface for external safety cover lock switch and alarm
7.6.4.1
Pin numbers
Connection
Plug type
Allocation
Meaning
Comment
24 V max.
9-pin
D-sub minia-
ture
Pin 1: S1
Pin 2: n.c.
Pin 3: n.c.
Pin 4: n.c.
Pin 5: DG
Pin 6: S2
Pin 7: n.c.
Pin 8: n.c.
Pin 9: DS
Buzzer No. 1
Not occupied
Not occupied
GND
Ext. safety cover unlock 0 V
Buzzer No. 2
Not occupied
Not occupied
Ext. safety cover unlock 24 V
External alarm
Battery link cover
External alarm
Battery link cover
Table 7.14:
Interface for external safety cover lock and alarm
ATTENTION:
The buzzer built-in voltage may amount to 24 V DC. The contact is in an open state of rest
and is separated galvanically. (That is, there is no electronic contact between the MediaClave and external
power supply. This is realised by relays or opto-electronic parts.) The forced opening cover only operates
with a battery and a maximum 24 V DC voltage.
ATTENTION:
Never open the vessel lid when operating the instrument.
ATTENTION:
The forced opening of the safety cover is not galvanically separated and must only be op-
erated with a battery.
Summary of Contents for MediaClave
Page 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Page 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Page 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Page 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Page 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Page 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Page 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Page 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Page 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Page 74: ...MediaClave overview INTEGRA Biosciences Page 68...