User Manual 3.4
INTEGRA Biosciences
Page 33
ENGLISH
OR
Press
CANCEL
to reject all modifications and return to the
PROGRAM
menu
11. Press
START
to start the selected program (see section 4.3.2.9)
4.3.2.9
Starting and running programs
1.
Open the safety cover and vessel lid
2.
Press
AUTOCLAVE
on the start display. The AUTOCLAVE PROG [#] screen appears
3.
Press
START
4.
Remove the cuvette from the vessel (see section 3.2)
5.
Insert the steel base plate (see section 3.2.2.3)
6.
Install the flexible temperature sensor (see section 3.2.2.2) and submerge it in the product
7.
Close the vessel lid
8.
Press
OK
to continue the cycle
OR
Press
END
to terminate the cycle
9.
Close the safety cover. The message
PREPARE COOLING
appears
•
All sensors are checked to ensure that they measure within the permitted temperature ranges. Sen-
sors are also checked for short-circuit, interruption and ice formation (< 2°C)
•
The pump runs for approximately five seconds to remove any trapped air
•
WATER CONTROL
appears on the screen. The water level is checked
•
PRESSURE TEST
appears on the screen
•
The sterilisation cycle is started
4.3.2.10
Flow of the AUTOCLAVE sterilisation cycle
A series of messages appears:
ATTENTION:
Ensure there is sufficient heating water in the vessel. If there is insufficient heating water
in the vessel, the following alarmed message appears –
TOO LITTLE HEATING WATER IN VESSEL
.
Press
END
to confirm this error. The cycle is terminated.
ATTENTION:
Pressing
END
results in the alarmed message
CYCLE IS BEING CONCLUDED
. The
MAIN
menu appears.
NOTE:
The safety cover lock ensures the safety cover is securely closed.
NOTE:
The heat exchanger is filled with water to prevent overheating. The actual and target temper-
atures are displayed in the upper right of the screen.
ATTENTION:
Press STEP to interrupt the cycle and proceed to the next step. The cycle cannot be in-
terrupted during cooling
NOTE:
A progress printout is printed every x minutes during the cycle (x refers to the selected print cycle,
see section 3.4.4.2).
NOTE:
The cycle progress is monitored and periodically updated on the GUI.
Summary of Contents for MediaClave
Page 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Page 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Page 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Page 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Page 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Page 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Page 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Page 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Page 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Page 74: ...MediaClave overview INTEGRA Biosciences Page 68...