User Manual 3.4
INTEGRA Biosciences
Page 53
ENGLISH
7.4.3
MEDIACLAVE program 14 (Without pressure test)
7.4.4
MEDIACLAVE program 15 (Chocolate Agar)
NOTE:
This program is capable of preparing 1–9 L of liquid. This program performs steam disinfection
but no sterilisation.
PROGRAMS 14
Range
Default
Sterilisation temperature:
70 - 105°C
105°C
Sterilisation time:
1 - 99 min.
20 minutes
Stirrer speed:
1,2
2 (150 rpm)
Dispensing temperature:
30 - 80°C
50°C
Stirrer speed:
1,2
1 (75 rpm)
Table 7.4:
MEDIACLAVE program 14
NOTE:
Blood is added to the sterilised medium and held for 10 minutes at 100°C (there is no sterilisation
of the added blood).
PROGRAMS 15
Range
Default
Sterilisation temperature:
70 - 122°C
121°C
Sterilisation time:
1 - 99 min.
20 minutes
Stirrer speed:
1,2
2 (150 rpm)
Cooling temperature:
30 - 80°C
50°C
Stirrer speed:
1,2
2 (150 rpm)
Boiling temperature:
70 - 122°C
100°C
Boiling time:
1 - 99 min.
10 minutes
Stirrer speed:
1,2
2 (150 rpm)
Dispensing temperature:
30 - 80°C
50°C
Stirrer speed:
1,2
1 (75 rpm)
Table 7.5:
MEDIACLAVE program 15
Summary of Contents for MediaClave
Page 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Page 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Page 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Page 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Page 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Page 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Page 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Page 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Page 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Page 74: ...MediaClave overview INTEGRA Biosciences Page 68...