Instructions
Specifications
External Power Supply:
Heated Lid: AC 10 V, 50-60 Hz
Heated Plate: AC 24 V, 50-60 Hz
Temperature Controller: AC 230 V/50 Hz or 115 V/60 Hz
Operating Conditions
Environmental operating temperature: 18-30°C (indoor use only)
Storage temperature: -5-50°C, humidity up to 80% RH
Humidity: 80% RH up to 30°C
Operating altitude: max. 2000 m (atmospheric pressure 800-1060 hPa)
Dimensions of Heated Plate with Lid
Width: 125.5 mm
Depth: 85.5 mm
Height: 25.5 mm
Weight: 210 g
Temperature Control Range
Heated Lid (Channel 1): Ambient temperature (min. 18°C) to +46°C
Heated Plate (Channel 2): Ambient temperature (min. 18°C) to +45°C
Recommended Temperatures before Adjustment
(see Section
Heated Lid (Channel 1): 42°C
Heated Plate (Channel 2): 37°C
Version (26th April 2017)
Heating System
7