Instructions
The cursor position is indicated with square brackets (”
[
xxx
]
”). You can move the cursor using the
”Left” and ”Right” buttons. If you want to select a parameter or a function, you need to press the
”Select” button. When you select a parameter, the square brackets will change to angle brackets
(”
<
xxx
>
”). Now you are able to change the value up or down. To confirm the changed value, you
must press the ”Confirm” button once more. When selecting a function, you will see the individual
function view, which is where you can change the parameters or access further functions.
Figure 12: Temperature Controller with Control Panel.
3.5
Temperature Adjustment in the Sample
The Heating System controls the temperature of the actively heated components. The temperature
in the sample will result from the interaction of all heated parts. Therefore, the temperature at the
position of the cells must be adjusted. An adjustment is recommended before starting the first ex-
periment. The temperature adjustment should be done for each chamber type, in order to control
temperatures with an absolute accuracy of less than 1°C.
Note!
The displayed temperatures are valid for the heated components only. The temperature near the
cells will be different!
The temperature adjustment must be renewed from time to time (at least once a year), especially if
one of the following conditions has been changed:
• Room temperature
• Air conditioning
• Chamber type or objective lens
• Humidification and gas flow
• Use of an XL-Incubator
• Use of an Objective Heater
Version (26th April 2017)
Heating System
25