Instructions
Supply Voltage Adjustments
The temperature controller must be adjusted to the corresponding voltage (230 V or 115 V). The
actual voltage setting may be read and adjusted at the integrated fuse box of the power cable
connector, located on the rear panel. In order to change these settings, it is necessary to pull out
the fuse holder from the casing. Replace the fuses to match the input voltage (1 A (T) for 230V /
50Hz or 2 A (T) for 115 V / 60 Hz). Rotate the grey fuse holder, and then reinsert it so that the
correct voltage is shown in the window. Snap the fuse box back into the housing. The ”On/Off”
switch is at the rear side of the controller. After selecting ”On”, the controller will automatically
start a self-test, and then the measured values will be shown.
2.4
Heated Lid
The Heated Lid (127.5 mm × 85.5 mm) provides excellent optical quality, and also allows for the use
of all standard microscopy techniques, including differential interference contrast (DIC). Due to the
height of the lid, we recommend using condensers with a working distance of
≥
26 mm.
The upper glass part of the Heated Lid is electrically heat controlled. The electrical cable (1.5 m) is
connected to the Temperature Controller. The Heated Lid fits exactly and securely onto the Heated
Plate.
For gas incubation, the Heated Lid is equipped with inlets for the gas flow and the humidity sensor.
Detailed information on the Gas mixer unit is given in the Gas Mixer instructions.
Figure 6: Heated Lid
2.5
Heated Plate
The Heated Plate provides the base for the Heating System. The lower part of the Heated Plate fits
into any microscope stage with a universal mounting frame for multiwell plates. The electrical cable
(1.5 m) is connected to the Temperature Controller.
20
Heating System
Version (26th April 2017)