MultiSystem 5060
Operating Instructions Manual
© Hydrotechnik GmbH • All rights reserved
Rev. 1.8/120307EN
Page 27 of 55
Example of a trigger recording
A 2 minute recording shall be started when the measured value for p2
falls below 50 bar and temperature T1 is higher than 30 °C. The re-
cording shall start 60 seconds before the trigger incident.
Required parameters
Storing time
2 min
Trigger 1
p2
Trigger type
LOWER
Trigger value
50.00
Pretrigger 50
%
Trigger link
AND
Trigger 2
T1
Trigger type
HIGHER
Trigger value
30
6.6 Submenu
“Device”
Configuration of the instrument is done with the device menu:
Language operation
language
Date current
date
Time current
time
ISDS
automatic sensor detection
Unit
select the unit system
CAN enable/disable
CAN
bus
Baud rate
set transmission speed for CAN data
HW-Filter hardware
filter
selection for each measuring channel
SW-Filter
software filter definition for each measuring channel
Press to display the 2
nd
page with device parameters:
Company
enter your company's name for the print-outs
Printer printer
selection
Format
print format selection
Keyboard
selection between STANDARD and COMFORT
Service
print a service report
RS232
select the speed of the RS 232 interface
Press to display the 3
rd
page with device parameters:
Ethernet m. select whether the optional ethernet module is con-
nected with the device
IP
enter the IP-adress in the ethernet network
Port
this is fix and displayed for information, only
Passwort
enter the password for the ethernet network
Additional functions
Setup: trouble-shooting informationen
Info: information on the software of the measuring instrument
OK: confirm and save changes
6.6.1
Select operation language
1. Open
function:
2. Select
language:
3. Confirm changes and leave function:
6.6.2 Enter
date
1. Open
function:
2. Enter
day:
3. Enter
month:
4. Enter year:
5. Confirm changes and leave function: