Assign 2
nd
scan rate
1. Highlight the line „Channels“ and press
.
2. Highlight the channels that shall be recorded with the 2
nd
scan
rate.
3. If „NO“ is displayed right of the channel, press
to select the
channel for recording „YES“.
4. Press
until „SLOW“ is displayed.
5. Press
again to assign the 2
nd
scan rate. A “*” will be displayed
beside the “YES” (see channels C5 and C6 in the screenshot).
6. Press
to save the settings.
Trigger function
You can use the trigger function to reduce the amount of stored data
by letting the instrument start the storing, when the "interesting mo-
ments" are coming. Here you can define up to two triggers.
Triggers are defined events that can start or stop a storing. You can
defined any channel as trigger, e.g. "if measured value at channel 2 is
higher than 10", use a timer function, or use a manual key pressure.
You can link two triggers logically, e.g. " if measured value at channel
2 is higher than 10 OR measured value at channel 6 is smaller than
100". The trigger will be started by the first of the two events.
1. Highlight the function "Trigger 1" –
.
2. Select a measuring channel, KEY (trigger is started by key pres-
sure) or TIMER (storing starts at a certain time) –
.
Definition of a measuring channel as trigger
3. Open the function "Type trigger" and highlight the desired option –
:
•
Greater: actuation when trigger value is exceeded
•
Lower: actuation when trigger value is fallen below
•
Rising: actuation when trigger value is fallen below for more
than 5% and then exceeded, "rising edge"
•
Falling: actuation when trigger value is exceeded for more than
5% and then fallen below, "falling edge"
4.
enter trigger value
.
Definition of a trigger time
3. Enter the date of the trigger time –
.
4. Enter the time value of the trigger time –
.
Define pretrigger
When a pretrigger is defined, the storing starts before the trigger event
has happened. The percentage defined as pretrigger (see above) is
used to store measured values before the trigger event.
5. Select the percentage of the pretrigger –
.
Trigger link
You can link trigger 1 with a second trigger:
6. Highlight the function "Trigger link" –
:
•
NONE: trigger 2 is disabled
•
AND: trigger 1 and trigger 2 must happen
•
OR: trigger 1 or trigger 2 must happen
•
START/STOP: trigger 1 starts the storing, trigger 2 stops it
7. Define trigger type and trigger value of trigger 2 like described in
the items 3. and 4.