Error messages / troubleshooting
FAM 25/45/60/75/95
en(us)
Page 120/160
BEWA FAM25-95 3402054f en(us) 2013-06-13
2013-06-13
To switch off the error messages, press the
button. If the fault indicator lamp is
flashing, error messages are present.
To show error messages again, switch to Menu 1 or Menu 2 and then return to the
display window.
1. CS+pump switched off. CS fault. Start 10 new attempts with Reset!
Cause:
The CS supplies 10 unusable measured values.
The CS and the charge pump are switched off for 15 seconds
after each unsuccessful attempt, after which a new
measurement is carried out.
After an additional 10 unusable measured values, the CS
switches off and can be switched back on only by actuating
the Reset key.
The reasons for incorrect measurements could include: no
usable values or one of the errors 16, 17, 30, 31.
Remedy:
Perform the following steps:
Step
Description
1. Message:
1. CS-Zuförderp. Abgeschal. CS1xxxStörung. Start 10 new
attempts with Reset!
2.
Actuate ESC key and reset failure with the F3 key.
3.
The measurement starts once again; automatic mode is not
interrupted.