Error messages / troubleshooting
FAM 15/30/50/70 ATEX
en(us)
Page 53 / 96
BeWa FAM15-70 Atex 3193049h en(us) 2013-06-19
2013-06-19
[H8] V-pump temperature
Step
Description
1.
Malfunction:
"V-pump temperature"
Cause:
The vacuum pump temperature is too high
2.
Check the fill level in the vacuum pump. If necessary, fill up with
the correct vacuum pump oil
3.
Press the RESET button to acknowledge/reset the fault.
4.
Does the malfunction recur after the vacuum pump has started?
-> a. Yes
-> proceed to Step 6.
5.
-> b. No
-> proceed to Step 9.
6.
The vacuum pump or the controller in the control cabinet is
defective.
7.
Repair or replace the vacuum pump or the defective parts of the
controller.
9.
The fault is repaired.