58
6.6.
Angiography image operation (optional)
1.
The first procedure is same as ‘6.4 General Operation: procedure 2 ~ 7’.
•
On the angio mode, function of auto tracking is additionally available while scanning.
2.
Capture image and check image quality (in angiography mode).
(1)
Alignment and focus.
①
Move body with joystick and detect disc with disc tracking box. And Check the scan region box.
②
Start OCT scan by pressing scan start icon (
).
③
Move body with joystick slowly until disc is clearly recognized on the screen.
(2)
Capturing angiography image
①
Optimize OCT signal by pressing optimize icon on the screen (
) or optimize button on the body
(A).
②
For enhancing OCT signal, move position of reference mirror by pressing arrow of REF.M icon
(
) if needed.
③
If signal is still weak, change the Signal Level (
).
④
To prevent signal degradation and scanning errors that caused by focus shifting, use Auto tracking
mode by pressing the Disc Tracking icon(
) at the bottom if needed.
⑤
Press the button on joystick to capture image.