HOCT-1/1F
6.9.2.
Cleaning
1.
Cleaning objective lens and anterior segment adopter
(1)
Cover the objective lens with lens cap to protect the lens from external pollution.
(2)
Use blower for removing dust on the surface of lens.
(3)
Any contaminants on the objective lens will affect the measurement. Wipe them using soft cotton swab or
lens cleaning paper moistened with alcohol.
(4)
Be careful not to use the wrong tools, so as not to damage the surface of the lens.
(5)
When the Objective Lens Clean option in Setup mode is ON, the light is turned on to facilitate cleaning of the
Objective Lens.
(6)
If anterior segment adapter is used for patients with any infectious disease, be sure to clean the Anterior
segment adapter with cotton swab moistened with alcohol to prevent secondary infection.
2.
System exterior
(1)
Keep system exterior clean with soft cloth. For severe stains, wipe with a soft cloth with neutral detergent
diluted with water. Do not use organic solutions such as thinner or benzene.
(2)
Wipe the touch screen with dry soft cloth. Do not use sponge or cloth soaked with large amount of liquid.
(3)
Do not press hard or place magnetic objects near the touch screen.
3.
Part of patient contact
(1)
Wipe the headrest and the chinrest with a clean cotton swab or gauze. For severe stains, use a soft cloth
with alcohol.
(2)
Remove a single sheet of chinrest paper if the chinrest paper is used.
4.
Others
(1)
Cover device with dustcover for unused storage for a long time.
(2)
Clean headrest and chinrest with alcohol before sending device to authorized agent or Huvitz for
maintenance.
(3)
The rubber ring inserted to conceal the wires may be out during use. It can be used either by inserting it
again or by removing it.
Objective lens cap
Blower
Rubber ring
Summary of Contents for HOCT-1
Page 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1 1F USER MANUAL...
Page 77: ...HOCT 1 1F 3 Select the TRANSFER icon to send the report to the DICOM server...
Page 109: ...HOCT 1 1F 11 Selecting FULL Screen icon shows the current Bscan image in full screen...
Page 118: ...118 9 Selecting FULL Screen icon shows the current Bscan image in full screen...
Page 131: ...HOCT 1 1F Vessel Faz...
Page 149: ...HOCT 1 1F 8 3 Drawings of System...