6.9.3.
Replacement of consumables and fuse
1.
Replacing chinrest paper
(1)
Pull out two fixing pins from chinrest.
(2)
Put a new chinrest paper on the chinrest.
(3)
Insert two fixing pins into the chinrest paper hole.
(4)
Attach the chinrest paper to the chinrest.
2.
Replacing fuse
(1)
Ensure the power switch of device off (O).
(2)
Remove power cable from inlet.
(3)
Pull out fuse holder in the inlet with a tweezers.
(4)
Replace two new fuses in the fuse holder. Be sure to check the fuse specification for the replacement (250V
T 3.15AL).
(5)
Insert fuse holder into the inlet.
6.9.4.
Calibration
Huvitz recommends to calibrate the system once a year.
Contact to the HUVITZ’s service technicians or other authorized experts.
!
NOTE