HOCT-1/1F
2. Basic Display for an angiography
No
Name
Function
1
Screen setting
-
Single: display angio information about only single eye.
-
Both: display angio information about both of eyes.
-
Compare: compare angio data from two different days.
-
Progression: It is easy to check the progress by comparing the eyes taken
over several days.
-
Panorama: stitch several map and make one of big detail angio image.
2
Basic/Detail Display
Choose one between Basic angio map and Detail angio map.
3
Angio image
Select the angiogram in the combo box.
-
Superficial: Its signal depth is from ILM(0um) to IPL(0um).
-
Deep: Its signal depth is from IPL (0um) to OPL(0um).
-
Outer: Its signal depth is from OPL(0um) to BRM(0um).
-
Choriocapillary: Its signal depth is from BRM(15um) to BRM(45um).
-
Retina: Its signal depth is from ILM(0um) to OPL(0um).
-
Custom: Its signal depth is customized by user. As user set the depth range,
angiogram is also changed
4
5
6
7
8
Enface,
Thickness,
Depth coded map
Select the image in the combo box.
-
Enface: It represent Enface image.
-
Thickness: It represent Thickness map.
-
Depth coded map: Choriocapillary, Deep, Outer map is overlapped in this
image.
9
Layer and depth info
Inform the layer&depth of the displayed image. And User make the custom
angiogram by setting the layer&depth with this tool bar.
10
B-Scan
Acquired B-Scan with angiogram. And Tomography can overlap with a red
color for a blood probability.
11
Sensitivity slider
Control blood vessel sensitivity with this slider.
The SOCT algorithm calculates the decorrelation value for each pixel in the B-scan by comparing the OCT signal
intensity variations across the B-scans in each set. Static tissue locations, without flow, exhibit little variation in OCT
signal intensity over the repeated B-scans; Therefore, the decorrelation values would be low. Tissue locations with flow
(for example inside a flowing vessel), show large variations in OCT signal intensity over the repeated B-scans. At these
pixel locations, the decorrelation values would be high, indicating the presence of flow. The Angiogram image is a
❻
❽
⓫
❶
❷
❸
❺
❼
❿
❹
❾
Summary of Contents for HOCT-1
Page 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1 1F USER MANUAL...
Page 77: ...HOCT 1 1F 3 Select the TRANSFER icon to send the report to the DICOM server...
Page 109: ...HOCT 1 1F 11 Selecting FULL Screen icon shows the current Bscan image in full screen...
Page 118: ...118 9 Selecting FULL Screen icon shows the current Bscan image in full screen...
Page 131: ...HOCT 1 1F Vessel Faz...
Page 149: ...HOCT 1 1F 8 3 Drawings of System...