(2)
Check if there are asymmetric areas as right picture from the measured images.
-
If there are some, re-boot the device and measure again.
(3)
Check if there are any blurriness or smear as below picture from the scanned images.
-
If blurriness or smear appears, clean the finger prints or stain on the objective lens.
5.
Check OCT Scan
(1)
Select OCT Scan in MESURE Mode.
(2)
Select Macular 3D at Scan Pattern and change the OCT Sensitivity to Normal as below picture and then turn
the OCT Enface option to On.
NG