Sonicaid FM800 Encore Series Instructions for Use
107
16.4
116B
Maternal blood pressure error codes
Message
Description
User action
ARTEFACT
Irregular pulse or
excessive motion.
Make sure patient keeps still
during measurement.
AIR LEAK
Reading terminated
because target pressure
was not reached within 60
seconds.
The cuff is loose or not connected
properly.
OR The cuff bladder or hose has
a leak.
OR The hose is not properly
connected to FM800E.
OVER
PRESSURE
Cuff over-pressurised.
Measurement terminated
automatically.
Make sure the patient does not
press the cuff, or bend her arm too
much.
MALFUNCTION
No reading reported.
Check cuff and hose, then try
another measurement.
If the problem persists, note the
error message, and contact
Huntleigh Healthcare Ltd or their
representatives.