Sonicaid FM800 Encore Series Instructions for Use
101
249B
Stopping Sonicaid CARE analysis
Stop printing. FM800E prints the analysis results.
250B
Turning Sonicaid CARE analysis off
> SETUP
> ANTEPARTUM ANALYSIS
> ANALYSIS OFF
The analysis also defaults to OFF when FM800E is switched off, then on again.
13.6
108B
Sonicaid CARE analysis report
When the analysis is stopped, the printer produces a report of the analysis results at the
end of the trace. The report shows:
Values for the calculated parameters
When the Dawes/Redman criteria were first met
Whether the Dawes/Redman criteria were met at the time the analysis was stopped
Abnormalities
251B
Reasons for not meeting the criteria
If the criteria were not met when the analysis was stopped, the reasons are given as
coded numbers alongside the CRITERIA NOT MET message:
Code Reason
1
Basal heart rate outside normal range
2
Large decelerations
3
No episodes of high variation
4
No movements and fewer than 3 accelerations
5
Baseline fitting is uncertain
6
Short-term variation is less than 3ms
7
Possible error at the end of the record
8
Deceleration at the end of the record
9
High-frequency sinusoidal rhythm
10
Suspected sinusoidal rhythm
11
Long-term variation in high episodes below acceptable level
12
No accelerations