70
CALIBRATION AND MAINTENANCE
Figure 314.
Select the desired calibration procedure and follow the
on-screen instructions.
Figure 310.
Calibration Check
A quick calibration check can be performed on the
balancer using a calibration weight.
From the main balance screen, touch the “Tools” button.
Figure 311.
Touch the “Quick Calibration Check” button.
Figure 312.
Screw in the calibration weight and clamp a large cone
or collet to the shaft. Lower the hood and press start to
begin the check.
Figure 313.
Rotate the shaft to the 12:00 position until the yellow dot
on the screen turns green to complete the check.