35 | 82
OWNERS MANUAL | C-FLEX | D0579-B DECEMBER 2019
NOTICE
–
Using the Laser ON and Laser OFF buttons in the software will require toggling of the
system key switch to resume emission, all laser lines will be turned OFF when the key is toggled.
See section 5
for instructions on how to stop and start emission.
7.
Stop the emission on all laser lines. Be sure the shutters on each laser are open, as well as the system
shutter.
8.
Place the aperture pin hole in the C-FLEX aperture. The collar is intended to be used for positioning the
pinhole in the center of the aperture. The pin hole alignment tool can be used in place of the aperture
mounted pin hole. It is not necessary to fix the pin hole with screws but it is advised to orient the article
number markings in the same direction for all measurements.
9.
Place a fast optical power meter 10 cm from the aperture mounted pin hole and secure it in place.
Moving the detector can affect the measurement.
10.
Resume the emission of the laser line in the position farthest from the aperture. For example, the red
laser line shown below in position 6.
This is considered the ‘reference laser’.
11.
Measure the power through the aperture mounted pin hole.